5000 Process Moisture Analyzer

Features

The versatile AMETEK 5000 is a highly sensitive moisture analyzer for process applications where the continuous, reliable measurement of moisture in vapors and gases is required. The 5000 analyzer measures moisture on-line by monitoring the vibrational frequency change of a hygroscopically sensitized quartz crystal that is exposed alternately to wet sample and dry reference gas for 30 second periods. The reference gas is sample gas passed through a molecular sieve dryer to remove virtually all its moisture content.

TYPICAL APPLICATIONS

  • Catalytic Reforming
  • Moisture in Olefins
  • Moisture in Natural Gas

OTHER APPLICATIONS
The 5000 provides similar solutions to problems in a variety of other applications, including:

  • Petroleum Refining – reforming; alkylation; LPG; light hydrocarbons; isomerization
  • Natural Gas – cryogenic extraction; transmission stations; storage/distribution; LNG; production
  • Petrochemicals – cracked gas; propylene; butadiene; ethylene
  • Chemicals – fluorocarbon gases; vinyl chloride; vinyl fluoride; refrigerants; reactor gas blankets; methyl chloride
  • Enhanced Oil Recovery – CO2 pipelines
  • Electronics – doping gases; dielectric gases; soldering furnace atmospheres; brazing furnace atmospheres; blanketing gases
  • Metals – annealing furnace atmospheres

Industrial

  • Oil & Gas
  • Chemical & Petrochemical

Description

Specifications

Ranges Calibrated 0 to 1000 ppm by volume (ppmv). Display c ontinues to provide trend information above 1000 ppmv. Output capability in lb/mmscf and dewpoitn (requires sample line pressure as analog input; single point systems only)
Outputs  80-character vacuum fluorescent display. Two 4-to-20 mA analog signals with completely user-settable span; one analog output is autoranging. RS485 bidirectional serial port
Alarms (3)  System alarm; Concentration alert; Range or calibration alert
Sensitivity  0.1 ppmv or 0.5% of reading, whichever is greater
Accuracy  0 to 20 ppm; 1 ppm; greater than 20 ppm; 5% of reading
Reproducibility  0.2 ppmv or 1% of reading, whichever is greater
Operating Pressure  103 kPa (15 psi) gauge. Allowable Inlet Pressure: 207 to 690 kPa (30 to 100 psi) gauge
Sample Gas Temperature 0º to 100ºC (32º to 212ºF)
Gas Flow Requirements All systems are available with choice of nominal AC voltages
100± 10%, 50/60 Hz
115± 10%, 50/60 Hz
230± 10%, 50/60 Hz
Field Unit: 175W
Controler: 95-230 V ± 10%, 47-63 Hz, 75W
Maximum Separation 600m (2000 ft.) between field unit and controller
Ambient Temperature Limits Field unit: -20º to 50ºC (-4º to 122ºF) without heat; -40º to 50ºC (-8º to 122ºF) with optional electric heat
Controller: -10º to 50ºC (14º to 122ºF)
Moisture Generator Value Exact values stated with analyzer; Nominal: 20 ppmv; Low option: 3 ppmv
Certificate UL/CSA General Safety Requirements (general purpose)
UL/CSA Class I, Division 2 NEMA 4X, Groups A, B, C, D, T4
UL/CSA Class I, Division 1, Groups B, C, D, T; IP65
ATEX II 2 G EEx d IIC T6 or EEx de IIC T; IP65
Russian GOST 1ExdIICT6X
Russian GOST Pattern Approval
Complies with all relevant European directives